设为首页 | 加入收藏
Chen Liyi.Polarographic determination of orthophosphate in seawater[J].Acta Oceanologica Sinica,1988,(1):68-73
Polarographic determination of orthophosphate in seawater
Polarographic determination of orthophosphate in seawater
投稿时间:1986-08-10  修订日期:1987-06-09
DOI:
中文关键词:  
英文关键词:
基金项目:
作者单位
Chen Liyi Department of Chemistry, Jishou University, Jishou, China 
摘要点击次数: 828
全文下载次数: 636
中文摘要:
      In the system HCl-Sb (Ⅲ)-ammonium molybdatc-acetone-butanone, PO43-, Sb (Ⅲ) and Mo (VI) form ternary heteropoly acid. This ternary heteropoly acid anion is adsorbed on the dropping mercury electrode and reduced quickly into heteropoly blue. The process of electrochemical reaction will produce a sensitive polarographic wave with peak potential -0.42V (vs S.C.E.). The detection limit is 6×10-8 mol/l. Variable coefficients are 10% and 3% for 0.1 μmol/l and 1 μmol/l respectively.
This method is sensitive, accurate and extremely rapid for analysis of seawater.
英文摘要:
      In the system HCl-Sb (Ⅲ)-ammonium molybdatc-acetone-butanone, PO43-, Sb (Ⅲ) and Mo (VI) form ternary heteropoly acid. This ternary heteropoly acid anion is adsorbed on the dropping mercury electrode and reduced quickly into heteropoly blue. The process of electrochemical reaction will produce a sensitive polarographic wave with peak potential -0.42V (vs S.C.E.). The detection limit is 6×10-8 mol/l. Variable coefficients are 10% and 3% for 0.1 μmol/l and 1 μmol/l respectively.
This method is sensitive, accurate and extremely rapid for analysis of seawater.
查看全文   查看/发表评论  下载PDF阅读器