设为首页 | 加入收藏
ZHANG RONGKUN,LUO YINGHUA.RESEARCH ON DETERMINATION OF DISSOLVED OXYGEN IN SEAWATER[J].Acta Oceanologica Sinica,1982,(1):62-70
RESEARCH ON DETERMINATION OF DISSOLVED OXYGEN IN SEAWATER
RESEARCH ON DETERMINATION OF DISSOLVED OXYGEN IN SEAWATER
投稿时间:1979-10-16  
DOI:
中文关键词:  
英文关键词:
基金项目:
作者单位
张荣坤 Xiamen University 
罗颖华 Xiamen University 
摘要点击次数: 752
全文下载次数: 545
中文摘要:
      Based on the principle of laminar flow-constant potential-chronocoulometry, this article presents a new method of determining the dissolved oxygen in seawater with considerable success. A new model of electrochemical sensor has been designed for this purpose.
The sensor has the advantage of overcoming hysteresis and "outskirts effect", thus increasing its stability and prolonging its use to 1-2 years.
In connection with this new method a new model SY-1 integrating microcoulorneter with digital display is designed.
It is a fast (a sample/2 min) and accurate (±1-1.5%) method, suitable for use in the laboratory on board a ship.
英文摘要:
      Based on the principle of laminar flow-constant potential-chronocoulometry, this article presents a new method of determining the dissolved oxygen in seawater with considerable success. A new model of electrochemical sensor has been designed for this purpose.
The sensor has the advantage of overcoming hysteresis and "outskirts effect", thus increasing its stability and prolonging its use to 1-2 years.
In connection with this new method a new model SY-1 integrating microcoulorneter with digital display is designed.
It is a fast (a sample/2 min) and accurate (±1-1.5%) method, suitable for use in the laboratory on board a ship.